Post Offer Free
Wuhan Huajin Co., Ltd.  

Gold Index: 6547

You are here: home  > Automatic Offset Calibration Function Microwave Power Detector Universal

Automatic Offset Calibration Function Microwave Power Detector Universal 

Place of Origin: Guangdong, China (Mainland) 
inquire
Add to My Favorites
HiSupplier Escrow
Share |

Product Detail

Model No.: AIA-1

Microwave Power Meter Available In Automatic Offset Cal

Microwave Power Meter Available In Automatic Offset Calibration Function

 

Product Overview:

 

This product is composed of microwave power meter main unit and serial power sensor, adopting embedded computer and embedded operating system integrally with application of microwave diode detection technology, digital signal processing technology and multidimensional compensation technology, which offer the instrument such characteristics as broad frequency band, large dynamic power range, sensor serialization, high precision, quick measurement and analysis and convenient operation. It is mainly used for the measurement and metering of average power, peak power and pulse envelope power of microwave signal, and also can be used for the measurement of high-power pulse modulation signal and narrow pulse modulation signal.

 

Main Features:

  1.  With accurate power measurement function of continuous wave signal, single sensor in dynamic range of 85 dB; 
  2.  Available in measurement and analysis functions of various envelope parameters of microwave and millimeter wave pulse modulation signal, such as: peak power, pulse power, average power, overshoot, rise time, fall time, top amplitude, bottom amplitude, pulse width, pulse period, duty ratio, closing time and pulse repetition frequency; 
  3.  Offering extended measurement function of high-power pulse if provided with high-power attenuator or high-power directional coupler to offer; 
  4.  Available in automatic offset calibration function; 
  5.  Available in automatic capture measurement function; 

 

Multiple measurement modes to meet different user’s measurement requirements

 

Available in continuous wave measurement, peak value measurement and CCDF statistical measurement, three measurement modes.

 

This instrument becomes a traditional continuous wave power meter if provide with continuous wave power sensor, which can measure the power of microwave and millimeter wave and continuous wave signals accurately, with single sensor in dynamic power measurement range up to 85dBm (-65dBm∼+20dBm). 

 

Typical Applications:


AV2436 microwave power meter is a universal high-performance measurement instrument for peak microwave power, mainly used for measuring and metering average power, peak power and pulse envelope power of microwave signal, and also can be used for the measurement of high-power pulse modulation signal and narrow pulse modulation signal.

 

In continuous wave mode, it is a universal traditional microwave power meter;

In peak value measurement mode, through different timebase settings, the instrument can automatically measure and analyze the peak value power, pulse power, average power, overshoot, rise time, fall time, top amplitude, bottom amplitude, pulse width, pulse period, duty ratio, closing time and pulse repetition frequency, etc. multiple envelope parameters of microwave and millimeter wave pulse modulation signal. 

 

General Technical Specifications:  


Main technical specifications of main unit

 

bigPhoto
 Mr. Angel

Tel: 86-27-86659761
Contact to this supplier

Related Search

Find more related products in following catalogs on Hisupplier.com

Company Info

Wuhan Huajin Co., Ltd. [China (Mainland)]


Business Type:Manufacturer
City: Wuhan
Province/State: Hubei
Country/Region: China (Mainland)

You May Like:

Product (141)


Frequency range

10MHz∼40GHz (depends on power sensor )

Power range

-65dBm∼+20dBm (depends on power sensor )

Maximum display resolution

Logarithmic mode: 0.001dB
Linear mode: bit 4

Offset range

±100.00dB

Rise time

≤13ns

Timebase range

10ns/Div∼1h/Div

Minimum detectable pulse width

50ns

Pulse width accuracy

50ns∼500ns: ≤13ns
501ns∼1s: ≤2%

Maximum detectable pulse repetition frequency

20MHz

Pulse repetition frequency accuracy

1Hz∼20MHz: ≤2%

Internal trigger level range

-20dBm∼+20dBm

Calibration
source

Frequency

1GHz±10MHz

0dBm Power accuracy