Product Overview:
This product is composed of microwave power meter main unit and serial power sensor, adopting embedded computer and embedded operating system integrally with application of microwave diode detection technology, digital signal processing technology and multidimensional compensation technology, which offer the instrument such characteristics as broad frequency band, large dynamic power range, sensor serialization, high precision, quick measurement and analysis and convenient operation. It is mainly used for the measurement and metering of average power, peak power and pulse envelope power of microwave signal, and also can be used for the measurement of high-power pulse modulation signal and narrow pulse modulation signal.
Main Features:
Multiple measurement modes to meet different user’s measurement requirements
Available in continuous wave measurement, peak value measurement and CCDF statistical measurement, three measurement modes.
This instrument becomes a traditional continuous wave power meter if provide with continuous wave power sensor, which can measure the power of microwave and millimeter wave and continuous wave signals accurately, with single sensor in dynamic power measurement range up to 85dBm (-65dBm∼+20dBm).
Typical Applications:
AV2436 microwave power meter is a universal high-performance measurement instrument for peak microwave power, mainly used for measuring and metering average power, peak power and pulse envelope power of microwave signal, and also can be used for the measurement of high-power pulse modulation signal and narrow pulse modulation signal.
In continuous wave mode, it is a universal traditional microwave power meter;
In peak value measurement mode, through different timebase settings, the instrument can automatically measure and analyze the peak value power, pulse power, average power, overshoot, rise time, fall time, top amplitude, bottom amplitude, pulse width, pulse period, duty ratio, closing time and pulse repetition frequency, etc. multiple envelope parameters of microwave and millimeter wave pulse modulation signal.
General Technical Specifications:
Main technical specifications of main unit
Frequency range |
10MHz∼40GHz (depends on power sensor ) |
|
Power range |
-65dBm∼+20dBm (depends on power sensor ) |
|
Maximum display resolution |
Logarithmic mode: 0.001dB |
|
Offset range |
±100.00dB |
|
Rise time |
≤13ns |
|
Timebase range |
10ns/Div∼1h/Div |
|
Minimum detectable pulse width |
50ns |
|
Pulse width accuracy |
50ns∼500ns: ≤13ns |
|
Maximum detectable pulse repetition frequency |
20MHz |
|
Pulse repetition frequency accuracy |
1Hz∼20MHz: ≤2% |
|
Internal trigger level range |
-20dBm∼+20dBm |
|
Calibration |
Frequency |
1GHz±10MHz |
0dBm Power accuracy |