Place of Origin: | Guangdong, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
AV3985 Millimeter-wave Noise Figure Analyzer
Product Overview:
AV3985 Millimeter-wave Noise Figure Analyzer adopts several high and new technologies such as wideband track pre-selection, wideband low-noise millimeter-wave amplification, multi-functional integrated miniature design, high-speed sweep measurement, embedded PC hardware platform, Windows 2000 software platform, and Chinese interface, which is easy of realizing high-precision measurement of noise figure and gain in linear or quasi-linear network of microwave and millimeter-wave at 10 MHz ~ 40 GHz.
It is widely used in the measurement of noise figure and gain of microwave and millimeter-wave amplifier, mixer, attenuator, etc. components and devices and front end linear circuit part of receiver.
In comparison with the foreign testing program for noise figure at wave band of 8mm achieved through spread spectrum, AV 3985 has been improved substantially in respect of frequency precision, noise figure precision and gain testing range, etc., representing the highest level of technology for testing noise figure at wave band of 8mm both at home and abroad.
Main Features:
Wide frequency coverage, compact system configuration
With wide frequency coverage, this instrument can measure the noise figure and gain within the frequency range from 10MHz to 40 GHz, compact design and good repeatability of system could minimize the error and uncertainty resulted from system setup. The integrated model replaces the complex testing system divided into microwave and millimeter-wave.
Perfect loss compensation
Loss compensation function can compensate the loss in channels before and after DUT in the form of fixing, table and combination, which greatly facilitates testing on the microwave and millimeter-wave tube core requiring deembedding.
Abundant peripheral interfaces
Multiple peripheral interfaces, such as GP-IB, RS232, LAN, USB, VGA and universal computer keyboard and intelligent mouse interface, greatly facilitate re-setup of various applications and systems.
Technical Specifications
Frequency Range |
10MHz~40GHz |
|
Measurement Bandwidth |
4MHz, 2MHz, 1MHz, 400kHz, 200kHz, 100kHz |
|
Frequency Tuning Accuracy |
4MHz: ± (Frequency Reference Error+100kHz) 10MHz~3GHz |
|
Noise Figure Measurement Range |
0~30dB (Noise Source ENR:14dB~16dB) |
|
Noise Figure Measurement Uncertainty |
±<0.15dB (Noise Source ENR:14dB~16dB) |
|
Gain Measurement Range |
-20dB ~ +40dB |
|
Gain Measurement Uncertainty |
±<0.17dB |
|
This Unit Noise Figure |
<7.5dB 10MHz~500MHz |
|
Frequency Tuning Accuracy |
4MHz: ± (Frequency Reference Error +100kHz) 10MHz~3GHz Wuhan Huajin Co., Ltd. [China (Mainland)]